Welcome: Shanghai source pulse Biological Technology Co., Ltd
Language: Chinese ∷  English
cart

Cooling circulating water device UC series

  • UC052F split chiller
UC052F split chiller

UC052F split chiller

  • Product description: Split chiller, consisting of indoor and outdoor units. The outdoor unit of this series of products is composed of a refrigeration compressor, a heat exhaust fan, and a condenser, which discharge all t

UC052F split chiller

Composed of indoor and outdoor units. The outdoor unit of this series of products is composed of a refrigeration compressor, a heat exhaust fan, and a condenser, which discharge all the heat generated by refrigeration to the outside. At the same time, the main noise source is also outdoors. The indoor unit consists of a water pump, water tank, and evaporator, without generating heat and with low noise. The split design concept greatly reduces the noise level of the equipment and improves the comfort of use. Simultaneously, it enables the equipment to output greater cooling power.
1. Environmentally friendly chillers, suitable for laboratory equipment with high requirements for environmental comfort and noise;
2. LCD display screen, displaying the actual temperature, set temperature, and return difference simultaneously, making it clear at a glance;
3. High performance circulating water pump. The multiple purification water quality design of the circulating water system prevents impurities from entering the circulating pump system, resulting in a longer service life of the pump;
4. Advanced PID control technology with high temperature control accuracy;
5. Stable and reliable system: high-performance and high-quality spare parts with stable quality and reliable performance;
6. Multiple safety protection: temperature protection, water pressure protection, overheat and overcurrent protection, and simultaneous warning with light and sound during alarm;
7. The refrigeration series adopts imported compressors and silent fans to reduce the noise during operation, making it suitable for applications in laboratory environments;
8. When giving an alarm, both light and sound are used to remind, and various alarm codes are directly displayed to directly determine the error point;

9. The maximum and minimum temperature settings of the machine can be limited.

Product parameters:
Product model: UC052F
Working temperature range [° C]: 5... 35
Temperature stability [± K]: ± 1.0
Temperature display: LED
Display accuracy [° C]: 0.1
Heating power [W]:  /
Cooling capacity (20 ° C) [W]: 5200
Pump head [m]:  /
Maximum pump flow rate [L/min]: 28
Tank volume [L]: 60
External dimensions (WxDxH) [mm]: indoor: 445x565x890, outdoor: 825x310x790
Instrument fixation method: floor mounted
Allow continuous operation [%]: 100
Allowable ambient temperature [° C]: 5.32
Allowable relative humidity [%]: 80
Pump connection interface [mm]: RC1/2 (4-point internal tooth) interface, which can be matched with various specifications of pipelines
Weight [KG]:  /
Packaging size (WxDxH) [mm]:  /
Order number: 101025002000